Sterilization is an essential step in the process of producing sterile  medical devices. To guarantee sterility, the process of sterilization  must be validated. Because there is no direct way to measure sterility,  the techniques applied to validate the sterilization process are based  on statistical principles. Steam sterilization is the most frequently  applied sterilization method worldwide and can be validated either by  indicators (chemical or biological) or physical measurements. The steam  sterilization conditions are described in the literature. Starting from  these conditions, criteria for the validation of steam sterilization are  derived and can be described in terms of physical parameters. Physical  validation of steam sterilization appears to be an adequate and  efficient validation method that could be considered as an alternative  for indicator validation. Moreover, physical validation can be used for  effective troubleshooting in steam sterilizing processes.
Subscribe to:
Post Comments (Atom)
 
Hi,
ReplyDeleteSterilization is an essential step in the process of producing sterile medical devices. It is the most frequently applied sterilization method worldwide and can be validated either by indicators chemical or biological or physical measurements. Thanks a lot...
Instrument Sterilization